Model Number |
FT-600A |
1. Specification
1.1 Measure principle: ISE
1.2 Measurement method: direct testing
1.3 Sampling mode: manual/auto
1.4 Calibration mode: automatic two-point calibration
1.5 Testing item: K+, Na+, Cl-, Ca++, pH
1.6
Measurement range and resolution
|
Item
|
Measurement range
|
Resolution
|
|
K+
|
0.50 mmol/L -15.0 mmol/L
|
0.01 mmol/L
|
|
Na+
|
30.0 mmol/L -200.0 mmol/L
|
0.1 mmol/L
|
|
Cl-
|
30.0 mmol/L -200.0 mmol/L
|
0.1 mmol/L
|
|
Ca++
|
0.10 mmol/L -5.00 mmol/L
|
0.01 mmol/L
|
|
pH
|
6.00-9.00
|
0.01
|
1.7 Sample: serum, plasma, whole blood, diluted urine (except Ca++ and pH)
1.8 Throughput: ≥90 samples/h
1.9 Sample volume: ≤150μL
2. Parameter
|
Item
|
Accuracy(V)
|
CV value
|
Linearity (∣Δ%∣)
|
stability(ΔD)
|
Carry over(Δξ)
|
|
K+
|
≤0.3mmol/L
|
≤1.0%
|
≤3.0%
|
≤2.0%
|
≤1.5%
|
|
Na+
|
≤3.0mmol/L
|
≤1.0%
|
≤3.0%
|
≤2.0%
|
≤1.5%
|
|
Cl-
|
≤3.0mmol/L
|
≤1.0%
|
≤3.0%
|
≤2.0%
|
≤1.5%
|
|
Ca++
|
≤0.3mmol/L
|
≤1.5%
|
≤5.0%
|
≤2.0%
|
≤1.5%
|
|
pH
|
——
|
≤1.0%
|
——
|
——
|
——
|
3. Working environment
a) Power supply: ~220V±22V;50Hz±1Hz;
b) Temperature: 5℃~40℃;
c) Relative humidity: ≤80%;
d) Keep away from electromagnetic interference
e) Avoid direct highlight
f) Well grounding
4. Electric parameter
a) Power supply: ~220V±22V;50Hz±1Hz
b) Input power: 65VA
c) Fuse: BGXP-1A,Φ5×20
5. Function parameter
a) Operation: by English menu
b) Function: inside on-line clock, parameter set-up, maintenance, Q.C., trouble-shooting
c) Display: 320×240 LCD(CCFL)
d) Printer: internal printer(width:57mm)
e) Data storage: 3000 sample data and control sample data of 3 groups of recent 31 days
f) Interface: RS-232C
6 Size and weight
a) Size (H×W×D): 40cm×36cm×40cm
b) Weight: about 8kg
|